For Your Safety; Intended Use; Configuration; Activating The Device - Dräger X-am 1100 Notice D'utilisation

Masquer les pouces Voir aussi pour X-am 1100:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

For your Safety

Strictly follow the Instructions for Use
Any use of the device requires full understanding and strict
observation of these instructions. The device is only to be
used for the purposes specified here.
Maintenance
The maintenance intervals and measures specified in the
1)
Technical Handbook
as well as the specifications in the
Instructions for Use of the used DrägerSensors
carefully observed.
Repair of the instrument may only be carried out by trained
service personnel.
Accessories
Only the accessories specified in the order list in the
1)
Technical Handbook
may be used.
Safe coupling with electrical devices
Devices which are not mentioned in these Instructions for
1)
Use or in the Technical Handbook
can only be coupled
electronically after consultation with the manufacturers or
an expert.
Use in areas subject to explosion hazards
Devices or components for use in explosion-hazard areas
which have been tested and approved according to
national, European or international Explosion Protection
Regulations may be used only under the conditions
explicitly specified in the approval and with consideration
of the relevant legal regulations. The equipment or
components may not be modified in any manner. The use
of faulty or incomplete parts is forbidden. The appropriate
regulations must be observed at all times when carrying
out repairs on these devices or components.
Repair of the instrument may only be carried out by trained
service personnel according to Dräger Safety Service
Procedure.
Liability for proper function or damage
The liability for the proper function of the device is
irrevocably transferred to the owner or operator to the
extent that the device was serviced or repaired by
personnel not employed or authorized by Dräger Safety or
when the device was used in a manner not conforming to
its intended use.
Dräger Safety cannot be held responsible for damage
caused by non-compliance with the recommendations
given above.
The warranty and liability provisions of the terms of sale
and delivery of Dräger Safety are likewise not modified by
the recommendations given above.
____________
1) The Technical Handbook, Instructions for Use of the used sensors and
the PC software CC Vision for Dräger X-am 1100, 1700 and 2000 are
included on CD.
6

Intended Use

Portable gas measuring device for the continuous
monitoring of the concentration of several gases in the
ambient air in the working area.
Dräger X-am 1100, Dräger X-am 1700:
independent measurement of four gases.
Dräger X-am 2000, depending on the type of device:
independent measurement of one up to four gases.
1)
must be
What is What
1
2
2
3
7
6
4
0
5
1
Gas entry
2
Alarm LED
3
Buzzer
4
[OK] key
5
Power pack
6
[ + ] key
7
Display
Special symbols:
Fault
Warning
Display peak value
Show TWA
Show STEL
Bump test mode
Fresh air calibration

Configuration

To configure a device with standard configuration to meet
individual requirements, connect the device via the USB
infrared cable (order no. 83 17 409) or the E-Cal system
with a PC. The PC software "CC Vision" is used for
configuration.
– Changing the configuration: see the Technical Handbook
Standard gas configuration:
Measuring
range*
Alarm A1*
- setpoint
- can be
acknowledged
- self-latching
Alarm A2*
- setpoint
8
- can be
acknowledged
- self-latching
9
Standard device configuration:
Dräger
10
Bump test mode*
Fresh air calibr.*
11
Live sign*
12
13
14
Disable*
8
IR interface
LEL factor*
9
Fastening clip
10 Type plate
* Deviating, customer-specific settings can be selected
11 Charging contacts
upon delivery. The current settings can be checked and
changed using the CC-Vision software.
12 Measured gas display
13 Measured value display
14 Special symbols

Activating the Device

Before using the device for the first time, insert the
supplied batteries or the NiMH power supplx unit T4 (order
1-key calibration
no. 83 18 704), see the Chapter "Replacing the
Singel gas calibration
Batteries".The Dräger X-am 2000 is ready for use.
Dräger X-am 1100 and Dräger X-am 1700 only
Password required
You also have to carry out the following activation
Battery 100 % full
sequence:
Battery 2/3 full
Press and hold the [ + ] key for approx. 3 seconds until
the countdown » 3 . 2 . 1 « shown in the display has
Battery 1/3 full
elapsed.
Battery empty
– The remaining operating time (not in the case of Dräger
X-am 2000) is displayed, for example » d 730 «
(remaining operating time 730 days).
– After that, the device is automatically switched off again.
You can switch it on again at any time by pressing the
[OK] key.
____________
1)
.
2) In the case of O
Ex
O
CO
H
2
[%LEL]
[vol. %]
[ppm]
[ppm]
0 to 100 0 to 25
0 to 2000 0 to 200
2)
20
19
30
10
yes
no
yes
yes
no
yes
no
no
40
23
60
20
no
no
no
no
yes
yes
yes
yes
X-am 1100 / X-am 1700 / X-am 2000
disable
disable
enable
not disabled
4.4
(4.4 % by vol. correspond to 100 %LEL)
A1 is the lower alarm setpoint: alarm if the value is too low.
2
S
2

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

X-am 1700X-am 2000

Table des Matières