QwikCheck™
Electronic Noise:
Sperm Quality Analyzer Guide Version 1.00
•
Reboot the QwikCheck system without a testing capillary in the chamber:
•
Turn the system OFF then back ON at the main switch on the rear panel.
•
Press the front panel ON/OFF key to begin Auto-Calibration and
Stabilization.
•
Call technical support if this message is displayed again. Prepare for technical
support by printing a copy of the SERVICE DATA:
•
Press the SERVICE key on the keypad to activate the SERVICE MENU
screen.
•
Select: PRINT SELF TEST DATA AND DEFAULT SETTINGS>SELF
TEST DATA.
•
Press ENTER
•
Ensure there is no testing capillary in the measurement compartment.
•
Remove the QwikCheck from sources of electronic noise (cell phones, etc.)
and vibrations (centrifuge).
•
Nettoyer le compartiment de mesure (référer vous à l'annexe) et, après
nettoyage:
•
Éteindre le système et le rallumer à partir du bouton principal situé à
l'arrière de l'instrument .
•
Presser le bouton ON/OFF au clavier sur la face avant et laisser les
procédures de stabilisation et d'auto-calibration se dérouler.
•
A partir du menu principal ( MAIN menu): Sélectionner TEST NEW PATIENT
et répéter le test.
•
Appeler votre founisseur si le message apparaît à nouveau. Pour preparer
l'intervention, veuillez procéder à l'impression des valmeurs du SERVICE
DATA:
•
Presser la touche SERVICE sur la clavier avant afin d'activer le menu
SERVICE.
•
Selectionner: PRINT SELF TEST DATA AND DEFAULT
SETTINGS>SELF TEST DATA.
•
Presser: ENTER