Applying The Cuff To The Patient - Omron HBP-1100 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 43
Applying the Cuff to the Patient
The device can be used on either
the right or left arm.
Wrap the cuff on a bare arm or
over thin clothing. Thick clothing
or a rolled up sleeve will cause
inaccurate blood pressure
measurements.
1. Make sure the cuff tube is not bent.
The cuff tube should be on the peripheral side.
2. Wrap the cuff so that the INDEX ARTERY "
the brachial artery.
The brachial artery is on the inner side of the patient's upper arm.
Make sure that the INDEX ARTERY "
the
pressure value. In this case, use the appropriate cuff size.
* Attach the cuff so that the bottom edge is 1 to 2 cm from the inner
side of the elbow joint.
* The cuff should be wrapped to a tightness that roughly allows two
fingers to be inserted under the cuff.
3. During measurement, keep the brachial artery on which the cuff
is wrapped at the same height as the right atrium of the heart.
" is within the range. If outside
range, there will be greater error in the blood
" is directly over
1 to 2 cm
15