6.2 Func-CYCL

Automatic recording of measurement values
In this mode, up to 10,000 records can be saved. To do this, select the item Func = CYCL in
the Configuration menu <SEt LoGG>. An indicator arrow ( ) now appears on the display
above logg. Automatic data recording at intervals of the set cycle time can be started by
button press. To do this, proceed as follows:
Data are deleted via the data archive.

7 Universal output

The universal output can be deactivated, used as a serial interface (factory setting:
<SEt Out> = SEr), or used as an analogue output. If not in use, we recommend deactivating
the output (<Set Out> = off) to keep battery consumption as low as possible.
Plug assignment
4: External supply (+5 V, 50 mA)
3: GND
4
1
2: TxD / RxD (3.3 V logic)
3
2
1: +U
DAC
52
Start measurement value recording:
A long press on the <Store> button in operating mode causes the
<Logg Run> prompt to appear. Confirm with
The indicator arrow above logg starts to flash > <
Stop measurement value recording:
A long press on the <Store> button during data recording causes
the <Logg Stop> prompt to appear. Confirm with
, analogue output
.
.
Only suitable
!
adapter
cables are
permitted