1 Specifications
on page 3
2 General information
on page 3
Section 1 Specifications
Specifications are subject to change without notice.
Specification
Measuring range
Response time
Repeatability/precision (0–20 µS/cm)
Precision (20–200,000 µS/cm)
Maximum cable length
Warranty
Specification
Linearity
Measuring range
Response time
1
Precision
Maximum cable length
Warranty
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.
1
Radio frequency fields in the 700–800 MHz range can cause inaccurate results.
Table 1 Contacting conductivity
Table 2 Inductive conductivity
Details
≥ 1.5 mS/cm: ±1% of reading; < 1.5 mS/cm: ±15 µS/cm
0–2000 mS/cm
0.5 seconds
> 500 µS/cm: ±0.5% of reading; < 500 µS/cm: ±5 µS/cm
200 to 2000 µS/cm: 61 m (200 ft); 2000 to 2,000,000 µS/cm: 91 m (300 ft)
1 year; 2 years (EU)
3 Installation
on page 5
Details
Cell constant 0.05: 0–100 µS/cm
Cell constant 0.1: 0–200 µS/cm
Cell constant 0.5: 0–1000 µS/cm
Cell constant 1: 0–2000 µS/cm
Cell constant 5: 0–10,000 µS/cm
Cell constant 10: 0–200,000 µS/cm
0.5 seconds
±0.1/0.1 µS/cm
±0.5% of reading
91 m (300 ft)
1 year; 2 years (EU)
English 3