Guidance and manufacture's declaration – electromagnetic immunity
The ri-thermo
®
gnetic environment specified below. The customer or the user of ri-thermo
fastPRObe Thermometer should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequen-
cy (50Hz/60Hz)
magnetic field IEC
61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
fastPRObe Thermometer is intended for use in the electroma-
IEC 60601
Test level
Con:
± 8 kV
Ais:
± 15 kV
N / A
N / A
N / A
30 A/m
Compliance
level
Con:
± 8 kV
Ais:
± 15 kV
N / A
N / A
N / A
30 A/m
Electromagnetic en-
vironment - guidance
Floors should be wood,
concrete or ceramic
tile. If floor are covered
with synthetic material,
the relative humidity
should be at least 30%.
The quality of the
supply voltage should
be that of a typical
business or hospital
environment.
The quality of the
supply voltage should
be that of a typical
business or hospital
environment.
The quality of the
supply voltage should
be that of a typical
business or hospital
environment.
Mains frequency ma-
gnetic fields should be
at a level charac-
teristic of a typical
location in a typical
commercial hospital
environment.
®