probe insert and about 30 seconds is placed next to the Thermometer.
The measurement check time for the water bath testing last approx. 60 seconds.
Attention!
We recommend a technical check in two-year intervals.
7.3
Standard
This thermometer conforms to all of the requirements established in
ASTM Standard E1104
ASTM Standard E1112
DIN EN ISO 80601-2-56
8.

Technical Data

Medical Equipment:
Electrical protection:
Electrical degree of protection:
Operating mode:
Dimensions:
Weight:
Battery
Temperature Measurement
Range:
Ambient Operating
Temperature Range:
Ambient Operating
Air pressure:
Storage and transport
conditions
Storage and transport
Air pressure
Laboratory Accuracy
Water Bath Accuracy (36°C to
42.0°C):
Direct measuring mode Accu-
racy (All measuring points)
Fast measuring mode Accuracy
(All measuring points)
Typical Measurement Times
(After position of the probe)
Probe Cover Material
46
Electronic thermometer
Electrical protection:
Type B
Continuous manual operation
205 x 70 x 70mm
400g with Batteries and Probe cover box
4x AA, LR6 Alkaline 1,5 V or
4x AA Mignon NiMH 1,2 V
34 °C - 43 °C (93,2 °F to 109 °F)
10 °C to 40 °C (50 °F to 104 °F)
at 10% - 85% non-condensing
700 to 1050 hPa
-20 to 55 °C (-4 °F to 131 °F),
at 10 - 95 % non-condensing
700 to 1050 hPa
Direct Measuring Mode ±0.1°C (±0.2°F)
Fast Measuring Mode ±0.3°C (±0.5°F)
± 0,1 °C (± 0,2 °F)
± 0,3 °C (± 0,5 °F)
Direct Measuring Mode
Oral Mode: 60 seconds
Axillary Mode: 60 seconds
Rectal Mode: 60 seconds
Fast Measuring Mode
Oral Mode: 12 seconds
Axillary Mode: 12 seconds
Rectal Mode: 12 seconds
HDPE