- We recommend to remove the batteries of ri-thermo
fastPRObe Thermometer before cleaning or disinfection.
- Take care when cleaning and disinfecting the Thermometer and
the probe, so that no liquid penetrates into the interior.
- Never place the ri-thermo
removable parts of the ri-thermo
(probe, cables) in liquids!
- The thermometer is shipped non-sterile. DO NOT use ethylene
oxide gas, heat, autoclave or any other harsh methods to
sterilize the unit.
- The devices are not meant to undergo machine-processed
maintenance and sterilization. This could lead to irreversible
damage!
- The probe cover is suitable for single use only!
The patient are not the intended Operator.
The device could be used by trained personnel.
Trained personnel is a doctor, nurse in hospitals, medical facilities,
clinics, doctors' offices.
1.6

User Responsibility

2.

Using the Device for the first time

2.1

Scope of Supply

Artikel- No. 1830
Artikel- No. 1831
34
Attention!
It is your responsibility:
- Before each use, the user must check the integrity and
completeness of the thermometer, the probe with spiral cable
and the probe cover. All components must be compatible with
each other.
Incompatible components can result in degraded performance.
- Check calibration of the device every two years.
- Never knowingly use a defective device.
- Immediately replace parts that are broken, worn, missing,
incomplete, damaged or contaminated.
- If the probe is damaged, it must be replaced immediately with
a new probe. The damaged probe must not be used any further!
- Contact the nearest factory approved service center should
repair or replacement become necessary.
- Further, the user of the device bears sole responsibility for any
malfunction that results from improper use, faulty maintenance,
improper repair, damage or alteration by anyone other than
Riester or authorized service personnel.
Any serious incident that has occurred in relation to the device
should be reported to the manufacturer and the competent
authority of the Member State in which the user and/or patient
is established.
ri-thermo
Probe cover box (20 pc / Box)
User Manual
4 AA Alkaline Batteries
ri-thermo
Probe cover box (20 pc / Box)
User Manual
4 AA Alkaline Batteries
fastPRObe Thermometer or
®
fastPRObe Thermometer
®
fastPRObe Thermometer Oral / Axillary
®
fastPRObe Thermometer Rectal
®
®

Publicité

Table des Matières
loading

Table des Matières