Description of the device
8
7
1. Colour display
3. Arrow button up
5. "RUN/STOP" button
7. "On/Off" button
9. Measuring funnel
11. Camera
13. Power socket
15. Battery compartment
GB
1
2
3
4
5
6
2. Function buttons "F1", "F2" and "F3"
4. "ENTER" button
6. Arrow button down
8. "ESC" button
9
10
11
12
13
14
15
10. Temperature and humidity sensor
12. USB connection
14. Tripod thread
Operating manual – Particle counter PC200

Technical data

Parameters
Model
Dimensions H x W x D
Weight
Particle counter
Channels (detectable particle sizes)
Flow rate
Counter mode
Counting efficiency
Coincidence loss
Zero count
Start delay
Sample inlet
Calibration
Light source
PC interfaces
Tripod thread
Storage conditions:
Display
Functions
Image format, resolution
Video format, resolution
Data storage
Energy
Battery
Operating time
Charging time
Automatic switch-off
Automatic screen switch-off
Temperature measurement
Temperature range
Temperature accuracy
Dew point temperature range
Dew point temperature accuracy
Wet-bulb temperature range
Wet-bulb temperature accuracy
Humidity measurement
Measuring range for humidity
Humidity accuracy
Values
PC200
240 mm x 75 mm x 57 mm
570 g
0.3 μm, 0.5 μm, 1.0 μm, 2.5 μm, 5.0 μm, 10.0 μm
2.83 l/min. (0.1 ft³/min.)
Controlled by internal pump
Cumulative, differential, concentration
50 % at 0.3 μm;
100 % for particles > 0.45 μm (according to ISO 21501)
5 %, 2 million particles per 28.3 litres
1 count/5 minutes (according to JIS B9921)
1 to 100 seconds
Isokinetic probe
Using a monodisperse latex particle
(PSL particle; compliant with NIST)
Laser class 3B, wavelength 780 nm, 90 mW
USB connection
1/4 inch – 20 UNC
-10 °C to +50 °C
< 95 % RH (non condensing)
2.8 inch colour LCD display with backlight
320 x 240 pixels
Maximum, minimum and average display,
measured value hold function, alarm function,
language selection, °C/°F switch,
photo or video recording
JPEG, 640 x 480 pixels
3GP, 320 x 240 pixels
5000 data records on internal Flash memory
(optional memory expansion with MicroSD card: max. 8 GB)
NiMH battery
approx. 4 hours of continuous operation
approx. 2 hours with an AC adapter
3 minutes, 15 minutes or 60 minutes
90 seconds, 2 minutes or 4 minutes
0 °C to 50 °C (32 °F to 122 °F)
±0.5 °C (0.9 °F) at 10 °C to 40 °C (50 °F to 104 °F)
±1.0 °C (1.8 °F) at other temperature ranges
-30 °C to 100 °C (-22 °F to 199 °F)
±0.5 °C (0.9 °F) at 10 °C to 40 °C (50 °F to 104 °F)
±1.0 °C (1.8 °F) at other temperature ranges
0 °C to 80 °C (32 °F to 176 °F)
±1.0 °C (1.8 °F)
0 % RH to 100 % RH
±3 % RH at 40 % to 60 %
±3.5 % RH at 20 % to 40 % and 60 % to 80 %
±5 % RH at 0 % to 20 % and 80 % to 100 %
B - 3