7. Troubleshooting
Problem
"Preventative Maintenance
Required. Call for Service"
message appears
"Replace Air Filter" message
appears
"Air Filter Missing.
Install Filter"
Lost USB storage device
with all cycle data
No Internet Connection
Possible Causes and Solutions
The annual maintenance should be done after 365 days
have elapsed since the initial installation or the last
service call was performed. Call for service.
• Replace HEPA filter. See section 6.3.
• If filter was replaced and message continues to
appear, reset drying counter. See section 3.11.
• Check HEPA filter to ensure it is installed correctly
(with arrow in correct position).
If you need to copy all of the unit's cycle data to a new
USB storage device:
1. Insert the new USB device.
2. Select the USB icon.
3. Select cycle information.
4. From the cycle information screen, select the USB
icon at top right.
5. Select copy and all of the unit's cycle history will be
copied to the new device.
Select the connectivity icon to go to the connection
status screen.
• If a red X appears next to IP: You do not have
a network connection. Contact your network
administrator.
• If a red X appears next to Internet: You do not have an
Internet connection. Contact your Internet provider.