Table 5 pH and ORP sensor wiring – sensors with removable cables (continued)
Terminal
8-pin (J5)
1
2
3
4
5
6
7
8
2-pin (J4)
1
2
Sensor shield wires—Connect all sensor
ground/shield wires to the controller enclosure
grounding screws.
Notes:
Connector
8-pin (J5)
2-pin (J4)
Description
S7 double
shielded
Reference
Inner stranded
wire (silver)
Ground solution
Jumper 1–2 on J5
–V supply
Temp
Temp/–Circuit low
Active
Transparent
Outer stranded
wire (copper)
Table 6 5500 series dissolved oxygen sensor wiring
Pin no.
1
2
3
4
5
6
7
Temp/–Circuit low
8
1
2
Cable type (connector)
S7 single
shielded
Inner stranded
wire (silver)
Jumper 1–2 on J5
Transparent
Signal
Reference
–V supply
+5 V
–5 V
Temp
Shield
Active
AS9
MP4
Outer stranded
Outer stranded
wire (copper)
wire (copper)
Jumper 1–2 on J5
Jumper 1–2 on J5
Brown
White
Transparent
Transparent
(core)
(core)
Sensor wire
Red
Jumper to J4 pin 1
Green
Blue
White
Yellow
Black
Silver (two wires)
Jumper to J5 pin 2
English 7