IMPORTANT PARTS OF YOUR CONCENTRATOR

Please take time to familiarize yourself with your Drive DeVilbiss oxygen concentrator before operating .
1
2
A
6
Front View (Figure A)
1 .
Operating instructions
2 .
Power Switch
| = ON
O = OFF
3 .
Flow meter knob
4.
Flow meter
5.
Circuit breaker – resets the unit after electrical overload shutdown
6.
Oxygen outlet – oxygen is dispersed through this port
7 .
Normal Oxygen (green) light (see page 8)
8 .
Low Oxygen (yellow) light (see page 8)
9.
Red Service Required (red) light – when illuminated contact your
DeVilbiss provider
10 . Hour meter

Accessories

Transfiller Caddy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .DeVilbiss 525DD-650
High Flow (6-15 LPM) Bubble Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Salter Labs 7900 or equivalent
Low Flow (up to 6 LPM) Bubble Humidifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Salter Labs 7600 or equivalent
High Flow Nasal Cannula . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Salter Labs 1600HF or equivalent
There are many types of humidifiers, oxygen tubing and cannulas/masks that can be used with this device. Certain humidifiers and accessories may impair the device's
performance. A mask or any nasal cannula can be used with continuous flow delivery and may be sized according to your prescription as recommended by your homecare
provider who should also give you advice on the proper usage, maintenance and cleaning .
WARNING
When using the Transfiller Caddy with a Transfill device, always keep the system on a flat surface. Disassemble the system prior to moving.
NOTE– The device should only be used with bubble humidifiers that are designed for use with flows up to 10 liters per minute and 20 psi pressure.
NOTE– A maximum of 50 feet (15 meters) of crush-proof oxygen tubing, plus 7 feet (2.1 meters) of cannula, plus a bubble humidifier is allowed between the concentrator
and the patient.
NOTE– Your healthcare provider should verify the compatibility of the oxygen concentrator and all of the parts used to connect to the patient before use.
EN - 6
9
8
7
3
4
5
10
B
15
16
11
14
12
13
Back View (Figure B)
11 . Handgrip
12 . Exhaust
CAUTION
When device is used under extreme operating conditions, the
temperature near the exhaust vents on the bottom of the unit
may reach 138.2˚F (59˚C). Keep body parts a minimum of 30"
(76 .2 cm) away from this area .
13 . Power cord and/or IEC power connector .
14. Line cord strap
15. Air filter – prevents dirt, dust and lint from entering your unit.
16 . Auxiliary Oxygen Port: Your concentrator is equipped with an auxiliary
oxygen port that can be used to fill oxygen cylinders with an FDA-cleared
cylinder filling device that is designed to use oxygen from a concentrator
to fill a cylinder. The port is only for use with FDA-cleared filling devices
with compatible oxygen input specifications. Refer to the cylinder filling
device instruction guide for the oxygen input/output specifications,
connection and operating instructions .