DANGER
Oxygen causes rapid burning . Do not smoke while your oxygen concentrator is operating, or when you are near a person utilizing oxygen therapy .
Smoking during oxygen therapy is dangerous and is likely to result in facial burns or death . Do not allow smoking within the same room where the oxygen
concentrator or any oxygen carrying accessories are located .
If you intend to smoke, you must always turn the oxygen concentrator off, remove the cannula and leave the room where either the cannula or mask or
the oxygen concentrator is located . If unable to leave the room, you must wait 10 minutes after you have turned off the oxygen concentrator before
smoking .
Oxygen makes it easier for a fire to start and spread. Do not leave the nasal cannula or mask on bed coverings or chair cushions if the oxygen concentrator is
turned on but not in use. The oxygen will make the materials flammable. Turn the oxygen concentrator off when not in use to prevent oxygen enrichment.
Keep the oxygen concentrator and cannula at least 2 m (6.5 feet) from hot, sparking objects or naked sources of flame.
Open flames during oxygen therapy are dangerous and are likely to result in fire or death. Do not allow open flames within 2 m (6.5 feet) of the oxygen
concentrator or any oxygen carrying accessories .
Drive DeVilbiss oxygen concentrators are equipped with a fire mitigating outlet fitting that prevents propagation of fire into the unit.
WARNING
To avoid electric shock, do not plug the concentrator into an AC outlet if the concentrator cabinet is broken . Do not remove the concentrator cabinet . The
cabinet should only be removed by a qualified Drive DeVilbiss technician. Do not apply liquid directly to the cabinet or utilize any petroleum-based solvents or
cleaning agents .
Improper use of the power cord and plugs can cause a burn, fire or other electric shock hazards. Do not use the unit if the power cord is damaged.
Ensure the mains power cord is fully inserted into the concentrator connector (230 volt units) and the power cord plug is completely inserted into a fully
functioning AC wall outlet . Failure to do so may cause an electrical safety hazard .
The use of a fire stop device as part of the oxygen accessories is optional, and is not contraindicated for use with this device. If used with the oxygen
accessories, the fire stop device should be located as close to the patient as practicable.
Locate oxygen tubing and power supply cords to prevent tripping hazards and reduce the possibility of entanglement or strangulation .
Do not lubricate fittings, connections, tubing or other accessories of the oxygen concentrator to avoid the risk of fire and burns.
Do NOT use lubricants, oils or grease .
Before attempting any cleaning procedures, turn the unit "Off ."
Use only water-based lotions or salves that are oxygen-compatible before and during oxygen therapy . Never use petroleum or oil-based lotions or salves to
avoid the risk of fire and burns.
Use only spare parts recommended by the manufacturer to ensure proper function and to avoid the risk of fire and burns.
When using the Transfiller Caddy with a Transfill device, always keep the system on a flat surface. Disassemble the system prior to moving.
If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately to avoid harm .
Geriatric, pediatric or any other patient unable to communicate discomfort can require additional monitoring and/or a distributed alarm system to convey the
information about the discomfort and/or the medical urgency to the responsible caregiver to avoid harm .
Use of this device at an altitude above 5000 feet (1524 meters) or above a temperature of 95°F (35°C) or greater than 93% relative humidity may affect the
flow rate and the percentage of oxygen and consequently the quality of the therapy. Refer to specifications for details regarding parameters tested.
To ensure you receive the therapeutic amount of oxygen delivery according to your medical condition, the Oxygen Concentrator must:
be used only after one or more settings have been individually determined or prescribed for you at your specific activity levels.
be used with the specific combination of parts and accessories that are in line with the specification of the concentrator manufacturer and that were used
while your settings were determined .
Your delivery settings of the oxygen concentrator should be periodically reassessed for the effectiveness of therapy .
For your safety, the oxygen concentrator must be used according to the prescription determined by your physician .
Under certain circumstances, oxygen therapy can be hazardous . Seek medical advice before using an oxygen concentrator .
WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the
oxygen concentrator or MR medical devices . The device and accessories have not been evaluated for safety in an MR environment .
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security
systems (metal detectors) as it may cause unacceptable risk to the patient or damage to the oxygen concentrator . Some electromagnetic sources may not be
apparent, if you notice any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and
discontinue use . Contact your home care provider .
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME
SYSTEM for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high .
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation . If such use is necessary,
this equipment and the other equipment should be observed to verify that they are operating normally .
Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the oxygen concentrator, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment
could result .
CAUTION
It is very important to follow your oxygen prescription. Do not increase or decrease the flow of oxygen – consult your physician.
To prevent product damage, do not attempt to operate the unit without the air filter, or while the filter is still damp.
The surface temperature of the exhaust vents on the bottom of the unit may exceed 105.8˚F (41˚C) under certain conditions.
When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 138.2°F (59°C) . Keep
body parts a minimum of 30" (76 .2 cm) away from this area .
Use of harsh chemicals (including alcohol) is not recommended . If bactericidal cleaning is required, a non-alcohol based product should be used to avoid
inadvertent damage .
EN - 4