The NEW HOSPIVAC 350 suction equipment does not need maintenance or lubrication.
It is necessary to check functioning and instrument before every use.
Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been
damaged during previous use.
Connect cable to electrical network and turn switch on.
Close the aspiration outlet with your finger and with suction regulator in maximum vacuum position check
that the vacuum indicators reaches -90 kPa (-0.90 bar) maximum.
Rotate the knob from right to left and check the aspiration regulating control.
The vacuum indicator should go down -40 kPa (-0.40 bar).
Verify that loud noises are not present, these can indicate wrong functioning.
A protection fuse (F 1 x 4A 250V) reachable from exterior and it situated in the plug protects the instrument.
For use replacing, always check the type and the range indicated.
Fault type
1. No aspiration
2. No aspiration
3. The float doesn't close
4. The float doesn't close
5. Low suction
6. No aspiration due to
flow leakage of mucus
Faults 1 - 2 - 3 - 4 - 5 - 6
If the overfill security system it's activated, don't proceede with the liquid aspiration.
If the overfill security system doesn't work there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device,
in this case return the device to CA-MI technical service.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
CA-MI DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE
MAINTENANCE
Cause
Jar Cap badly screwed
down
Lid seal not in its seat
If the cap has been
washed, ensure that the
float is not partially
detached
The float it's covered by
dirty material
Foam inside the jar
Filter blocked
None of the remedies has
achieved the desired
results
PLEASE CONTACT CA-MI TECHNICAL SERVICE.
CHECKING, APPEARS TO BE TAMPERED.
Unscrewed the cap, then rescrew it correctly
Unscrew the cap and insert the seal properly
Insert the float into it's place
Unscrewed the cap, leave the and put in on
Fill the jar to 1/3 full of ordinary water
Contact the seller or CA-MI After-sales
Assistance Service
14
Solution
in its seat
autoclave
Replace filter