Tv 551/701 Pump With Standard Rack Controller; Water Cooling Kit Installation - Varian TV 551 Navigator Notice De Mode D'emploi

Masquer les pouces Voir aussi pour TV 551 Navigator:
Table des Matières

Publicité

TV 551/701 Pump with Standard Rack
Controller
When the TV 551/701 pumps are used with the
standard rack controller, it is necessary to utilize
the air cooling kit model 969-9314.
Fan specification:
− air flow: 25 l/s (51 CFM)
− input voltage: 120 Vac 50-60 Hz
− maximum power: 17 W
− dimensions: 119 x 119 mm (4.69 x 4.79 inches)
To fix the fan to the pump, position it ensuring that
the holes in the plate line up with those in the
pump base.
Insert the respective washers and screws and
tighten with a screwdriver.
Connect the fan to the controller.

WATER COOLING KIT INSTALLATION

The water cooling kit mod. 969-9347 or 969-9337
must be installed when the pump is used under
heavy load conditions or when air cooling is insuf-
ficient.
Model 969-9337
Model 969-9347
The kit is assembled as shown in the figure.
Two rubber or plastic hoses from the water supply
must be fitted to the two water cooling banjos.
68