1. Automatic means that when the controller
stops, the vent valve is opened with a delay de-
fined by window n. 126; when the controller
starts, the vent valve is immediately closed.
2. On command means that the vent valve is
opened or closed by means of window n. 122.
Description
Admitted
Values
Pump tem-
0 to 70
perature in
°C
Pump status
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Error code
Bit description:
see the following
figure
Cycle time
0 to 999999
in minutes
(zeroed by
the reset
command)
Cycle num-
0 to 9999
ber (zeroed
by the reset
command)
Pump life in
0 to 999999
hours (ze-
roed by the
reset com-
mand)
CRC
QE5XXXX
EPROM
(where "XXXX"
(QE)
are variable)
CRC Pa-
PA5XXXX
ram. (PA)
(where "XXXX"
are variable)
RS 485 ad-
0 to 31
dress
(default = 0)
Serial type
0 = RS 232
select
1 = RS 485
(default = 0)
7
6
TOO HIGH LOAD
SHORT CIRCUIT
OVERVOLTAGE
AUX FAIL
Window N. 206 Bit Description

INLET SCREEN INSTALLATION

The inlet screens mod. 969,9302, 969-9304 and
969-9316 prevents the blades of the pump from
being damaged by debris greater than 0.7 mm di-
ameter.
The inlet screen, however, does reduce the pump-
ing speed by about 10%.
65
TECHNICAL INFORMATION
5
4
3
2
1
0
NO CONNECTION
PUMP OVERTEMP.
CONTROLL. OVERTEMP.
POWER FAIL
87-900-934-01(G)