Air Cooling Kit Installation; Tv 1001 Sem With Navigator Controller; Tv 1001 Sem Pump With Standard Rack Controller - Varian TV 1001 SEM Mode D'emploi

Table des Matières

Publicité

The following figure shows the overall flange di-
mensions with the protection screen fitted on
pump with ISO 200F flange and pump with ISO
200K flange.

AIR COOLING KIT INSTALLATION

TV 1001 SEM with Navigator Controller

An air cooling kit (mod. 969-9297) is available to
improve the TV 1001 SEM cooling during heavy
operational conditions (optional).
Fan specifications:
3
air flow: 200m
/h
input voltage: 24 Vdc
dimensions: 127x127x38 mm (5 x 5 x 1.5 in.)
power: 4.7 W
The fan bracket is shaped so that it can be
mounted close to the pump.
To fix the fan to the TV 1001 SEM case execute
the following procedure (see the following figure):
1. Fix the fan to the suitable bracket by means of
the furnished screws;
2. Fix the bracket to the pump body between the
pump and the controller;
3. Connect the fan supply to the P4 connector of
the controller.
Turbo V
-
TV 1001 SEM Pump with Standard Rack
Controller
When the TV 1001 SEM pumps are used with the
standard rack controller, it is necessary to utilize
the air cooling kit model 969-9315.
36
TECHNICAL INFORMATION
87-900-973-01

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-8939969-8940969-8978

Table des Matières