1. Use a clean, soft cloth to remove loose debris from the end of the sensor. Rinse the sensor with
clean, warm water.
2. Soak the sensor for 2 to 3 minutes in a soap solution.
3. Use a soft bristle brush to scrub the entire measuring end of the sensor.
4. If debris remains, soak the measuring end of the sensor in a dilute acid solution such as < 5%
HCl for a maximum of 5 minutes.
5. Rinse the sensor with water and then return to the soap solution for 2 to 3 minutes.
6. Rinse the sensor with clean water.
Always calibrate the sensor after maintenance procedures.

Replacement parts and accessories

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Sensors
Description
2 electrode conductivity sensor K=0.01, ¾ inch NPT thread
2 electrode conductivity sensor K=0.1, ¾ inch NPT thread
2 electrode conductivity sensor K=1, ¾ inch NPT thread
2 electrode conductivity sensor K=0.01, ¾ inch NPT thread
2 electrode conductivity sensor K=0.01, for Yokogawa flow chamber
2 electrode conductivity sensor K=0.01, ¾ inch G thread
2 electrode conductivity sensor K=0.1, ¾ inch NPT thread
2 electrode conductivity sensor K=1, ¾ inch NPT thread
2 electrode conductivity sensor K=0.01, 1½ inch (38 mm) clamp
2 electrode conductivity sensor K=0.01, 1½ inch (38 mm) clamp with certificate of
conformity
2 electrode conductivity sensor K=0.01, 2 inch (51 mm) clamp
2 electrode conductivity sensor K=0.01, 2 inch (51 mm) clamp with certificate of
conformity
Cables
Description
Female connector 6+T with connection drawing
5 m cable and IP65 connector for 2 electrode conductivity sensor
10 m cable and IP65 connector for 2 electrode conductivity sensor
20 m cable and IP65 connector for 2 electrode conductivity sensor
Shielded 4 conductor cable (per meter)
30 m cable and IP65 connector for 2 electrode conductivity sensor
8 English
Item no.
08310=A=0000
08311=A=0000
08312=A=0000
08315=A=0000
08315=A=0002
08315=A=1111
08316=A=0000
08317=A=0000
08394=A=1500
08394=A=1511
08394=A=2000
08394=A=2011
Item no.
08319=A=0000
08319=A=0005
08319=A=0010
08319=A=0020
588800,29050
91010=A=0144

Publicité

Table des Matières
loading

Table des Matières