The motor is located under the bowl of the
machine (fig 1).
Before starting, open the lid, switch off the
centrifuge and unplug power cable
CAUTION: THE FOLLOWING
OPERATIONS MUST BE PERFORMED
BY A QUALIFIED TECHNICIAN
Step 1: Remove the front panel then
disconnect motor wires and tachometer plug.
Also cut the collar tie
Step 2: Extract the motor seal then loose
the 4 screws of the motor and remove it.
(Clean the silicone around seal for
refrigerated model)
Please pay attention to the wires
Notice Originale - Rev.1.4
PROCEDURE AP2.010
MOTOR REPLACEMENT
Front panel removed
Manuel Maintenance – Lisa / Loreena
Fig 1
1/ 2