AFi LISA Série Manuel De Maintenance page 70

Table des Matières

Publicité

The motor is located under the bowl of the
machine (fig 1).
Before starting, open the lid, switch off the
centrifuge and unplug power cable
CAUTION: THE FOLLOWING
OPERATIONS MUST BE PERFORMED
BY A QUALIFIED TECHNICIAN
Step 1: Remove the front panel then
disconnect motor wires and tachometer plug.
Also cut the collar tie
Step 2: Extract the motor seal then loose
the 4 screws of the motor and remove it.
(Clean the silicone around seal for
refrigerated model)
Please pay attention to the wires
Notice Originale - Rev.1.4
PROCEDURE AP2.010
MOTOR REPLACEMENT
Front panel removed
Manuel Maintenance – Lisa / Loreena
Fig 1
1/ 2