The imbalance setting allows the safety during
the operation of the centrifuge.
The machine automatically detects the
imbalance and stops immediately.
Before starting the procedure switch on the
machine and open the lid.
CAUTION: THE FOLLOWING
OPERATIONS MUST BE PERFORMED
BY A QUALIFIED TECHNICIAN
Step 1: remove inserts from buckets and
go to IMBALANCE setting (in the service
menu)
Step 2: place 15 grams inside a bucket
and run the corresponding program
Notice Originale - Rev.1.4
PROCEDURE AP2.001
IMBALANCE SETTING
Manuel Maintenance – Lisa / Loreena
Preparing the operations
Weight inside the buckets
1/ 2