J2 - Serial; Rs 232/Rs 485 Communication Description; Communication Format; Connector Examples - Varian Turbo-V 81-AG Notice De Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11
J2 – SERIAL

J2 - Serial

This is a 9 pin D-type serial input/output connector to control
the TV 81Nav via an RS 232 or RS 485 connection.
PIN N.
1
+5 V (OUT)
2
TX (RS232)
3
RX (RS232)
4
Not used
5
GND
6
A + (RS485)
7
Not Used
8
B - (RS485)
9
Not Used
A serial communication kit with a serial cable and the
T-Plus software is available (optional).
RS 232/RS 485 COMMUNICATION
DESCRIPTION
Both the RS 232 and the RS 485 interfaces are available on
connector J2.
The communication protocol is the same (see the structure
below) but only the RS 485 manages the address field.
Therefore, to enable the RS 485, it is necessary to select the
type of communication as well as the device address by
means of the T-Plus.

Communication Format

8 data bit
no parity
1 stop bit
baud rate: 600/1200/2400/4800/9600 programmable
SIGNAL NAME

Connector examples

PC-HOST
0
TURBO
CONTROLLER
1
TURBO
CONTROLLER
2
The communication port mating connector is supplied with the
RS 232 PCB (AMP/Cannon or equivalent 15-pin "D" type male
connector).
For example, the Transmit data signal from controller (pin 2)
must be connected to the host computer's receive data line
(pin 2) and vice versa. Consult the host computer's Instruction
manual for its serial port connections.
Varian cannot guarantee compliance with ECC regulations on
radiated emissions unless all external wiring is shielded, with
the shield terminated on the metal enclosure on the
0-subconnector. The cable should be secured to the connector
with screws.
Serial Communication Descriptions
This unit can communicate using two different protocols:
- protocol (old system) "letter"
- "Windows" protocol (new system)
These two protocols can be used as well with 232 or
485 media.
Please use "Windows" protocol for new development.
46/50
TECHNICAL INFORMATION
CNT
TV81 Nav
RS – 232 Connection
120
1/4 W
A +
B -
Gnd
6
A +
8
B -
5
Gnd
6
A +
8
B -
Gnd 5
120
1/4 W
RS – 485 Connection
NOTE
NOTE