Technical Information; Controller Description; Controller Specifications; Controller Outline - Varian Turbo-V 81-AG Notice De Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 11

CONTROLLER DESCRIPTION

Two different versions of the controller are available:
969-8995
Turbo-V 81-AG Navigator Controller, 24Vdc
969-8996
Turbo-V
100 ÷ 240Vac
The controller is a solid-state frequency converter with the
following features:
Drives the pumps of the 81-NAV family.
Powers the pump cooling fan.
Drives the vent valve.
Provides and acquires the pressure of the wide range gauge.
Remote I/Os compatible with the previous version.
Navigator default serial compatible with the previous
RS 232 and 485 version.
The dedicated controller is a solid-state frequency converter
which is driven by a single chip microcomputer and consists of
two PCBs which include power supply and 3-phase output,
analog and input/output section, microprocessor and digital
section. The power supply, together with the 3-phase output,
converts the single phase AC mains supply or 24Vdc supply
into a 3-phase, low voltage, medium frequency output which is
required to power the pump. The controller can be operated
by a remote host computer via the serial connection. A Win-
dows-based software is available (optional).

CONTROLLER SPECIFICATIONS

Input:
(969-8995)
Voltage
Power
Input:
(969-8996)
Voltage
Frequency
Power
Fuse (969-8995)
Fuse (969-8996)
Output:
(969-8995)
Voltage
Frequency
Power
Output:
(969-8996)
Voltage
Frequency
Power
Environment storage
Operating
81-AG
Navigator
Controller,
24 Vdc ±10%
80W
100 ÷ 240 Vac ±10%
50 to 60 Hz ±5%
180 VA
2 x T4 A 250V
2 x T2 A 250V
60 Vac
1350 Hz
50W
60 Vac
1350 Hz
70 W (ramp), 50 W( Normal and
Autotuning)
-20 °C to +70 °C
0 – 95%
+5 ÷ +45 °C
0 – 90% (Not condensing)
Compliance with
standards
Safety:
Radio Interference:
Harmonics:
Flicker:
ESD:
Radiated RF Immunity:
Burst:
Surge:
Conducted RF Immunity:
Radiated Magnetic Field:
DIPS:
Weight:
Max. Altitude
Internal use only
Pollution Degree
Installation Category
All connecting cables for Remote I/O and serial connections, if
longer than 3m, must be of the shielded type
The Controller must be positioned in such a way that the
mains cable can be easily disconnected.
When the TV 81 Navigator has been stored at a temperature
of less than 5°C, wait until the system has reached the above-
mentioned temperature.

CONTROLLER OUTLINE

The outline dimensions for the Turbo-V 81-Nav controllers are
shown in the following figure.
35/50

TECHNICAL INFORMATION

EN 61010 - 1 (2001)
EN 55011 (Class-A)
EN 61000-3-2,
EN 61000-3-3,
EN 61000-6-2 (2005)
EN 61000-4-2
EN 61000-4-3
EN 61000-4-4
EN 61000-4-5
EN 61000-4-6
EN 61000-4-8
EN 61000-4-11
500g (230V) 300g (24V)
3000m
2
II
NOTE
NOTE
NOTE
87-901-004-01 (B)

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

969-8995969-8996

Table des Matières