Connection J2 - Semi Specifications Connector; Connection J7 - Accessories And Options Interconnections - Varian Turbo-V 2000 ICE Notice De Mode D'emploi

Masquer les pouces Voir aussi pour Turbo-V 2000 ICE:
Table des Matières

Publicité

Connection J2
Semi Specifications Connector
MALE CONNECTOR
(ANELVA CONNECTOR)
8
5
7
14
6
J2
13
5
12
I/O
4
11
3
10
2
9
1
Semi Specifications Connector
PIN
1-9
Remote START/STOP S1 optically isolated
from
the
internal
permanently closed contact (relay contact,
transistor, etc.). When the contact closes the
turbopump starts at high speed and when the
contact opens, the turbopump is stopped.
With the remote mode operation selected, the
front panel push-button is inoperative.
3-11
START S1 signal 24 V, 60 mA, isolated relay
output (open collector).
The output is activated when the START
push-button on front panel is pressed or the
remote start is present or the function has
been requested by RS 232/422/485, until
NORMAL operation is reached.
4-12
Normal S1 signal 24 V 60 mA, isolated relay
output (open collector).
The output is activated when the rotational
speed of the pump is higher than the selected
high speed threshold.
5-13
BRAKING signal 24 V, 60 mA, isolated relay
output
(open
activated when the pump is in its deceleration
phase, with exception of an
condition.
6-14
FAULT signal 24 V, 60 mA, isolated relay
output (open collector).
The output is activated when a fault condition
is displayed on the front panel display.
7-15
Remote/Local signal 24 V, 60 mA, isolated
relay output (open collector).
The output is activated when the controller is
configured in Remote Mode.
∗ Relay outputs
GND
REMOTE/LOCAL (OPEN CO LLECTOR)
GND
FAILURE (OPEN COLLECTOR)
GND
BRAKING (OPEN COLLECTOR)
GND
NORMAL S1 (OPEN COLLECTOR)
GND
STARTING S1 (O PEN COLLECTO R)
GND
START S1
DESCRIPTION
circuit,
requires
collector).
The
output
overtemp
Connection J7
Accessories and Options Interconnections
The 120 Vac, 1 A maximum (independent of line
Voltage) output Voltage is present when the main
power switch is set to 1 position and after START push-
button is pressed and will remain present until a fault
condition is displayed on the front panel display or the
turbopump is stopped.
The connector J7 is for vent device, forepump
connector coils, pump cooling fan etc.
a
To make connections, remove the plug and wire the
pins (maximum wire size 18 AWG, 1 mm²) as indicated
in the figure to obtain the desired capability.
Forepump relay coil is an independent user supplied
item.
is
74
TECHNICAL INFORMATION
J7 connectors
NOTE
87-900-929-01 (H)