Note: The setting operations of other parameters (such as TIME, PATIENT,
RECORDING INTERVAL, POWER SAVING etc.) are the same with date
setting.
- Time: Time setting
- Recording: Time interval for recording data (SpO
& PR), there are five
2
options: "1s, 2s, 4s, 8s" and "OFF".
1) "1s": the least length of data record is set to 30 seconds, and the maximal
length for one record is limited to 1 hour. The total storage time is up to 48 hours.
2) "2s": the least length of data record is set to 60 seconds, and the maximal
length for one record is limited to 2 hours. The total storage time is up to 96
hours.
3) "4s": the least length of data record is set to 120 seconds, and the maximal
length for one record is limited to 4 hours. The total storage time is up to 192
hours.
4) "8s": the least length of data record is set to 240 seconds, and the maximal
length for one record is limited to 8 hours. The total storage time is up to 384
hours.
5) When the option is set to "OFF", the device will not store the measuring data.
- Power saving: power saving setting; two options: "on" and "off". The factory
default setting is "on". If the power saving mode is on during measurement,
and there is no key operation for 2 minutes, the screen display will be dim
for power saving. The display brightness will resume to normal condition by
pressing any key.
- Volume (optional): 3 levels setting: "1", "2" and "3", corresponding to the
speaker volume icons of "
","
" and "
" respectively. The speaker
volume icon indicates the volume of any sound generated by the device for
key clicks, over-limit alerting, pulse beep and audible probe-off indication etc..