31
4.4.1 Setting
On the menu screen, select "Setting" and then press "
" key for entering
into system setup screen. The setup screen is as shown in the following figures.
Figure 4-4 System Setup Screen
Operation Instructions:
- Patient: two options:"ADU
" for adult, and "PED
" for pediatric.
- SpO2 Lo-Limit: SpO2 low limit setting; range: 50%~99%, the step is 1%. The
factory default value for adult is 90%
and 95% for pediatric.
- PR Hi-Limit: High limit setting of pulse rate; range: 100~240bpm. From 100
to 150, the step is 1bpm, and from 150 to 240, the step is 5bpm. The factory
default value for adult is 120bpm and 160bpm for pediatric.
- PR Lo-Limit: Low limit setting of pulse rate; range: 30~99bpm, and the step
is 1bpm. The factory default value for adult is 50bpm and 60bpm for pediatric.
Note: When the SpO2 reading is lower then the preset alert setting or the
PR reading is higher or equal to the preset alert setting, then the over-limit
alert event will be activated, that's, the alert sound "bibibibi" occurs, and the
corresponding reading(s) blinks. When measured on pediatric, if the SpO2
reading is lower or equal to the preset alert setting for 10 seconds, then the
alert sound and blinking display will be activated.
- Date: Date setting
When cursor stays on the Year of the date, press "
" (Mode/OK) key to
active Year option, the cursor flashes on the Year of the date;
Press "
/
" (Navigation key) to adjust year.
Press "
" (Power/back) key or "
" (Mode/OK) key to confirm and exit
from date setting.
The procedures of adjusting Month value and Day value are the same with
Year adjustment.
Date Format: yy-mm-dd