Autoclaving

The Transferpette
(250 °F), 2 bar absolute (30 psi), with a holding time of at least 15
minutes according to DIN EN 285.
1.
Eject the pipette tips.
2.
Autoclave the complete pipette without any further disassembling.
3.
Allow the Transferpette
Note:
The effectiveness of the autoclaving must be verified by the user.
Maximum reliability is obtained with vacuum
sterilization. We recommend the use of sterilization bags.
Attention:
Prior to autoclaving, the volume
adjustment must be set on an avail-
able numbered volume (e.g., 11.25
or 11.26 but not between), with the
volume-change protection set fully
unlocked.
If the pipette is autoclaved frequently, the
pistons should be oiled with the supplied
silicone oil in order to preserve smooth
movement.
If necessary after autoclaving, tighten the
connection between the hand grip and the
manifold.

UV sterilization

The unit can withstand the usual output of a UV sterilization lamp. The
effects of the UV may cause some color change.
40
S -8/-12 is completely autoclavable at 121 °C
®
S -8/-12 to completely cool and dry.
®