Important - BRAND Transferpette S-8 Mode D'emploi

Masquer les pouces Voir aussi pour Transferpette S-8:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Note:
ISO 8655 prescribes rinsing the pipette tips once with the sample liquid
prior to the actual pipetting process.

Important!

Don't lay the instrument horizontal when the tips are filled. Liquid may enter
and contaminate the instrument. The instrument should be stored without
tips, placed upright in the supplied shelf/rack mount or bench top rack.
5.
Discharge sample
a) Place the pipette tips against the wall
of the vessel. Hold the pipette at an
angle of 30-45° relative to the con-
tainer wall.
b) Press the pipetting button slowly to
the first stop and hold it down.
For serum and liquids of high viscos-
ity or low surface tension, observe
adequate waiting time to improve
accuracy.
c) The blow-out stroke empties the tips
completely: Press the pipetting button
down to the second stop.
d) While doing this, wipe the pipette tips
against the wall of the container.
e) Remove the pipette tips from the con-
tainer wall and let the pipetting button
slide back.
6.
Ejecting the tips
Hold the manifold over a suitable disposal
container and press the tip ejection key to
the stop.
Pipetting
35