Technical Data - Bosch+Sohn boso-medicus Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

Technical data

Measurement principle: oscillometric
Measurement range: 40 to 280 mmHg
40 to 200 pulses/min
Cuff pressure: 0 to 320 mmHg
Display: LCD
Operating conditions:
ambient temperature: 10°C - 40°C
relative humidity:
10% - 85%
Transport and storage conditions:
temperature:
-5°C - +50°C
relative humidity: max. 85%
Power supply:
DC 6V (4 x 1.5 V round cell IEC LR 6, alkaline manganese)
Alternatively as an extra: power supply pack DC 6V,
order no.: 410-7-150
(polarity: outer MINUS, inner PLUS)
Battery check: indicator symbol in display
O +
O –
O
O
Weight: 0.6 kg
Classification: Protection class II ( )
Type BF ( )
Clinical trials: DIN 58130 testing procedure N2
'Comparative Blood Pressure
Measurements on the Left and Right Arm'
Results:
systematic measurements variance:
systolic blood pressure: -0.98 mmHg
diastolic blood pressure:-3.2 mmHg
empirical standards variance:
systolic blood pressure: ±5.96 mmHg
diastolic blood pressure:±5.35 mmHg
Tolerance in cuff measurement: ±3 mmHg
Tolerance error in pulse rate measurement: ±5%
Technical specifications subject to change without prior
notification.
59

Publicité

Table des Matières
loading

Table des Matières