3 Make Measurement

3.1 SpO
Measurement
2
Operation procedures:
1. Connect the SpO
with "SpO
". (Note: When disconnecting the connector, be sure to hold the head of
2
the connector firmly and pull).
2. The red blinking light inside the clip of the SpO
nection.
3. Insert one finger (index finger is preferred, the nail should be not too long) into the
clip of the probe according to the finger mark,as shown in figure 3.1.
4. The device will begin to take the measurement, and the measured result will be
displayed on the screen, as shown in figure 4.2.
Figure 3.1 demonstration for SpO
Safety instructions for SpO
 Long term use of the SpO
pain, especially for those with microcirculatory problems. It is recommended that
the probe should NOT be applied to the same place for over two hours, change the
measurement site periodically and when necessary.
 When the ambient temperature is over 35℃, please change the measuring site every
two hours; when the ambient temperature is over 37℃, please do NOT use the SpO
sensor, as using in high temperatures can cause burns.
 Do NOT place the SpO
 Do NOT put the SpO
blood pressure measurement may affect the SpO
 The device is calibrated to display functional oxygen saturation
 Do NOT allow the sensor cable to twist or bend.
 Check the SpO
sor.
 When the temperature of the SpO
probe to the connector on the upper-side of the device marked
2
probe
2
measurement
2
probe on the same place may result in discomfort or
2
probe on a finger with edema or fragile tissue.
2
probe and pressure cuff on the same limb, otherwise the
2
sensor and cable before use. Do NOT use a damaged SpO
2
ENGLISH
probe indicates a successful con-
2
2
sensor is abnormal, do not use it further.
2
measurement.
2
sen-
2