Sartorius YDK01 Mode D'emploi page 7

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Screw the appropriate adapter into the underside
of the bar frame base:
– Compensating disk for ME235S/P and ME254S
balances only
– Gasket for all other ME models only
– Other adapter as required (see previous page)
Now remove the following components from
the balance:
– weighing pan
– short adapter for BA and BP balances
– pan support for BA and BP balances
Place the frame in the weighing chamber. The
wedge-shaped opening at the top of the frame must
face the direction from which the sample holder
(sieve/glass plummet) will be placed into the frame.
Always use the metal platform to hold a beaker.
Please position this platform through the frame so it
rests evenly on the base of the weighing chamber.
If you are using a BA, BP, CPA, CP or ME balance,
turn the platform so it rests on the pins which are
far apart. For RC and MC balances, the platform
should rest on the pins which are closer together.