EasyClean100/150
Electrical connections EasyClean150 with Mettler-Toledo transmitters
Electrical connections
Transmitter X100 e
NOTICE
Bridge terminal 2 and 3 in the EasyClean 150.
Define rinse interval and rinse time according to the process
requirements on the transmitter X100 e.
Transmitter M300
NOTICE
Bridge terminal 2 and 3 in the EasyClean 150.
Set relay #1 (default) or #2, #3, #4 on the transmitter to
rinse contact.
Define rinse interval and rinse time according to the process
requirements on the transmitter M300.
Transmitter M700
NOTICE
Bridge terminal 2 and 3 in the EasyClean 150.
Set relays K3 (or K1 or K2) to rinse contact and NO (normal-
ly open). Define the rinse interval and rinse time according to
the process requirements.
Set input OK1 to HOLD (active 10–30 V) for the external hold
signal.
Transmitter Trb 8300
Define rinse interval and rinse time according to the process
requirements on the transmitter Trb 8300.
EasyClean100 and 150 are designed to work together with Mettler Toledo transmitters.
Please consult relevant instruction manuals of third party transmitters for proper installation.
EasyClean100/150
52 500 178
EasyClean 150
Wash
X100e
EasyClean 150
TB1(1/2 DIN)
or
M300
TB1B (1/4 DIN)
EasyClean 150
K3
M700
EasyClean 150
Wash
Trb 8300
© 11/14 Mettler-Toledo AG, CH-8606 Greifensee
Printed in Switzerland
23
Hold
TB2
Hold
OK1