Examine catheter lumen and arterial and venous extensions after
each treatment for damage.
To prevent accidents, assure the security of all caps and bloodline
connections prior to and between treatments.
Use only Luer Lock (threaded) Connectors with this catheter.
Make sure catheter lumen is thoroughly dry before attaching the
adapter.
Repeated over tightening of bloodlines, syringes, and caps will
reduce connector life and could lead to failure of the arterial venous
extension connector.
When cutting catheter to desired length, assure the lumen is cut
square and that the remaining catheter lumen is not damaged.
Attach a syringe containing heparin solution to the female luer of
extension adapter. Open extension clamps. Aspirate to insure that
no air will be forced into the patient. Inject heparin into each
catheter using quick bolus technique.
DIRECTIONS FOR ASSEMBLY:
STEP 1
A
B
1. Take apart female adapter by twisting parts (A) and (C) apart. The
silicone compression ring (B) should be found in part (A).
STEP 2
A
B
2. Slide adapter part (A) over tubing. Slide compression ring (B) over
tubing. Insert metal portion of adapter part (C) into the tubing with a
twisting motion, making sure the tubing is fully seated against part
(C).
STEP 3
A
3. Slide compression ring (B) toward end of tubing/adapter assembly.
STEP 4
4. Slide adapter part (A) toward end of tubing/adapter assembly and
twist adapter together firmly until there is no gap between adapter
parts (A) and (C). A gentle tug will assure proper assembly.
STEP 5
5. Attach a syringe containing heparin solution to the female luer of
extension adapter. Open extension clamps. Aspirate to insure that no
air will be forced into the patient. Inject heparin into each catheter
using quick bolus technique.
Note:
If using StatLock
the adapter will lie on the patient with alcohol. Remove the backing of the
one side of the StatLock
remove the remaining protective backing. Apply slight pressure on the
pad to assure adherence. Push the collar section of the adapter into the
receiving grooves of the StatLock
C
C
C
B
A
C
for catheter securement, clean the area where
®
pad and position on patient. Once positioned,
®
pad. Repeat for second adapter.
®
-2-