6.5
Quick stop function
7
Software operation
7.1
Centrifugation parameters
7.1.1
Input with the SELECT button
RPM indicator
AB1801en
[START/PULSE] button to end the centrifugation run.
4.
Release the
�� Ramp-down takes place with the set brake level. The brake level is
displayed.
When the rotor is at a standstill, the lid opens, an audible signal
sounds and the remaining number of run cycles (centrifugation
runs) is displayed.
Personnel:
Trained user
[STOP/OPEN] button twice.
Press the
�� Ramp-down with brake level "fast" (shortest ramp-down time) is
displayed and executed.
The number of centrifugation parameters that can be set
differs depending on whether the RPM indicator or the RCF
indicator is selected.
This chapter describes input of the centrifugation parame-
ters with the RPM indicator and RCF indicator selected,
one after the other.
The display returns to the previous values if no button is
pressed for 8 seconds after parameter selection or during
parameter entry. The parameters must then be entered
again.
1.
If required: Press the
[RCF] button to toggle between the two parameters
�� Press the
'RPM' ) and RCF ( '>RCF<' ).
RPM (
[SELECT] button.
2.
Press the
�� Runtime in
't/min' is displayed
[t] buttons to set the desired value.
3.
Use the
Adjustable from 1 to 99 minutes in 1 minute increments.
The parameters t/min and t/sec must be set to zero to set continuous
operation.
'--:--' is displayed.
��
[SELECT] button.
4.
Press the
't/sec' is displayed.
�� Runtime in
Rev.: 10 / 11.2023
[RCF] button to select the RPM indicator.
Software operation