Operation
1
2
7
6
Fig. 15: Rotor installation and
removal
1
Button
2
outer ring
3
Groove
4
Motor shaft
5
Driver
6
Marker
7
Fastening nut
Installing the rotor
6.3
Loading
Filling centrifuge tubes
20 / 39
1.
Open the lid.
2.
Press down the button (
pull the outer ring (
3.
Remove the rotor lid from the rotor.
3
4.
Using the supplied spanner, unscrew the fastening nut (
5.
Lift the rotor off the motor shaft (
4
5
Personnel:
Trained user
The lid is open.
1.
Place the rotor vertically on the motor shaft (
5  ) on the motor shaft ( 4  ) must be in the groove ( 3  ) of the
driver (
rotor. The orientation of the groove is marked on the rotor (
2.
Hand-tighten the fastening nut (
supplied spanner.
3.
Check that the rotor is firmly seated.
4.
Place the rotor lid on the rotor and press down on the handle until the
rotor lid engages.
WARNING
Risk of injury from contaminated sample material.
Contaminated sample material escapes from the sample
tube during centrifugation.
NOTICE
Damage to the device due to highly corrosive substances.
Highly corrosive substances may impair the mechanical
strength of rotors, buckets and accessories.
Rev.: 10 / 11.2023
1  ) in the rotor lid handle. At the same time,
2  ) of the handle upwards.
4  ).
7  ) on the motor shaft ( 4  ) using the
Use centrifuge tubes with special screw caps for haz-
ardous substances.
For risk group 3 and 4 materials, use a biosafety system
in addition to the sealable centrifuge tubes (see WHO's
'Laboratory Biosafety Manual').
Do not centrifuge highly corrosive substances.
7  ).
4  ) and press down. The
6  ).
AB1801en