Commissioning
4.2
Fastening the transport lock
1
2
Fig. 13: Transport lock
1
Spacer sleeves
2
Screws
5
Commissioning
5.1
Unpacking the centrifuge
16 / 39
Personnel:
Trained user
The lid is closed.
The mains cable is disconnected from the device.
1.
Tilt the device on the back of the device.
2.
Insert 2 spacer sleeves (
3.
Screw in 2 screws (
1
2
CAUTION
Danger of crushing due to parts falling out of the transport
packaging.
CAUTION
Risk of injury from lifting heavy loads.
NOTICE
Damage to the device due to improper lifting.
Personnel:
Trained user
1.
Open the box at the top.
2.
Remove the padding.
3.
Remove the device and accessories by lifting them up out of the box.
4.
Place the device on a stable and level surface.
Rev.: 10 / 11.2023
1  ).
2  ).
Keep the device balanced during the unpacking
process.
Only open the packaging at the points provided for this
purpose.
Provide an adequate number of helpers.
Chapter 3.1 'Technical data'
Note the weight. See ⮫ 
on page 9 .
Do not lift the centrifuge by the control panel or the
control panel holder.
AB1801en