TECHNICAL SPECIFICATIONS
• This Flocare
Infinity
pump in combination with the charger and the Flocare
®
EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions and immunity and are in compliance
with Directive 93/42/EEC.
0344
• IEC 601: Class II
• Microprocessor controlled
• Type BF Shock Protection (EN60601-1) degree of protection against electrical shock. No electrical connection to the patient.
Drop from any height of 90cm shall not damage pump operation.
• Charger:
input 100-240V AC / 50-60 Hz / 0.4A Max.
output 5 V DC 3 A
• Battery: internal rechargeable Lithium ion battery, 3.7 V DC 2500 mAh
• Battery capacity: ~ 24 hours at 125 ml/h for a new battery.
• IPX 5: EN60529 degree of protection against water entering the enclosure. Water jets from any direction shall have no effect
on the pump.
• Dimensions:
~ 140 x 95 x 35 mm
• Weight: approx. 392 g
• Accuracy flow rate: ±5,0% with appropriate Flocare
• Humidity:
• Temperature:
• Atmospheric pressure:
• Occlusion detection pressure:
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies from 0.5 to 1,5 ml
which relates to an air bubble with an approximate length in the tubing of 9 - 26 cm.
• The Flocare
Infinity
can safely be operated on commercial aircraft.
®
• The use of other accessories, chargers and cables than listed within this manual may result in increased emissions or decreased
immunity of the equipment of the Flocare
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service accor-
ding to the EMC information provided in the accompanying documents.
• Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equipment. If the Flocare
Infinity
pump is used adjacent to or stacked with other equipment, the pump should be observed to verify normal operation.
• In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number.
• At the end of their service lifetime, the pump and its electrical accessories (AC Adapter Charger and Data Download
Accessory Cable), should be disposed of according to local standards and regulations governing the disposal of electronic
waste (e-waste). The pump contains a rechargeable lithium-ion battery. When disposing of the pump, be sure to discard this
equipment in a manner consistent with local policy for expired battery operated equipment. Other accessories can be dispo-
sed of or recycled as standard non-hazardous waste.
403260087_r1.indd 96
403260087_r1.indd 96
Medical Device Directive: Class IIa
Infinity
®
Operation mode:
Storage:
Operation mode:
Storage and transportation:
Operation mode:
Storage and transportation:
Upstream occlusion:
Downstream occlusion:
Infinity
pump.
®
Infinity
Nurse Call are designed to EN 60601-1-2,
®
pump set
30% to 75% noncondensing
10% to 95 % noncondensing
+5°C to +40°C
-20°C to +65°C
70-106 kPa
50-106 kPa
-34 kPa (tolerance 21 kPa)
83 kPa (tolerane 21 kPa)
-96-
®
18/04/23 10:27