Télécharger Imprimer la page

Nutricia flocare Infinity Mode D'emploi page 100

Masquer les pouces Voir aussi pour flocare Infinity:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 49
Guidance and manufacturer's Declaration – Electromagnetic immunity
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The customer or the user
®
of the Flocare
Infinity
enteral feeding pump should assure that it is used in such an environment.
®
Immunity test
Power frequency (50/60
Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Recommended separation distances between portable and mobile RF communications equipment and the
The Flocare
Infinity
enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF disturbances are control-
®
led. The customer or the user of the Flocare
portable and mobile RF communications equipment (transmitters) and the Flocare
output power of the communications equipment.
Rated Maximum output power of
transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in Watts (W)
according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
403260087_r1.indd 98
403260087_r1.indd 98
IEC 60601 test level
Compliance level
30 A/m
30 A/m
Flocare
Infinity
®
Infinity
can help prevent electromagnetic interference by maintaining a minimum distance between
®
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
Electromagnetic environment – guidance
Power frequency magnetic fields should be at levels cha-
racteristic of a typical location in the typical commercial
or hospital environment.
enteral feeding pump
Infinity
as recommended below, according to the maximum
®
80 MHz to 800 MHz
d = 1.2
CD
P
0.12
0.38
1.2
3.8
12
-98-
800 MHz to 6 GHz
d = 2.3
CD
P
0.23
0.73
2.3
7.3
23
18/04/23 10:27
18/04/23 10:27

Publicité

loading

Produits Connexes pour Nutricia flocare Infinity