ON/OFF
START/STOP
The "hold" mode is used to temporarily stop the flow of feed:
• to change the feeding program (installed parameters),
• to change the feed container,
• to administer medication without switching the pump off,
• to silence an alarm and correct problems.
After 3 minutes a two tone audible alarm sounds and the message "PUSH STRT" appears in the display.
Press "START/STOP" to stop the alarm and to extend the hold mode by a further 3 minutes or press "START/STOP" twice (3
beeps will be heard) to resume programmed settings.
ON/OFF
FILL SET
Ensure the feeding set is not connected to the patients feeding tube when the pump is in the hold mode:
Press and hold the "FILL SET" key for 2 seconds to activate the "FILL SET" maneuver. Release the button as soon as the
pump generates a beep and starts pumping at a flow rate of approx. 700ml/h. During this maneuver the message "FILL SET"
appears in the display. The pump will automatically stop when the Flocare
feed. However this "FILL SET" maneuver can at anytime be stopped by pushing the "FILL SET" key a second time. The pump
will return to the hold mode when the "FILL SET" maneuver is complete or stopped.
403260087_r1.indd 88
403260087_r1.indd 88
SWITCHING "OFF"
Press the "ON/OFF" key and keep it pressed for 2 seconds. A continuous alarm will be heard and the
pump switches off.
The feeding program (installed parameters) and total volume administered since the last clea-
rance will be retained in the pump's memory.
If the pump shuts down due to low battery voltage, the memory will be retained for 24 hours.
"HOLD" MODE
To temporarily pause the pump, or switch into "hold" mode whilst operating, press the "START/
STOP" key once. Three beeps are heard and the run symbol disappears.
The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered volume (ml) are
retained.
FILLING THE PUMP SET
The FILL SET function offered by the Infinity pump is used to fill a (new) feeding set with nutrition (or
water as the case may be). When the FILL SET function runs all alarms are deactivated helping avoid
alarms being inadvertently triggered (e.g. air alarm).
The FILL SET function should only be used to fill an empty (air-filled) feeding set. Use of the FILL SET
function at any other time may cause the pump to incorrectly calibrate.
Pack Infinity
pump set is completely filled with
®
-88-
18/04/23 10:27
18/04/23 10:27