connect the pump.
• The Flocare
Infinity
®
instructions for use of the Flocare
• For good pump operation, the pump should be kept clean and free of debris like (dried) tube nutrition. It is recommended to clean
the pump with a wet cloth after every finished feeding regimen and specifically in between patient uses (if applicable).
• Detergent and disinfectant cleaning agents and wipes can damage plastic surfaces of medical devices when they are not
compatible with the surface material. Damaged surfaces may compromise the ability to decontaminate medical devices adequately
and / or may interfere with device function and affect their integrity and performance. Ensure cleaning solutions, detergents and
disinfectant wipes are compatible with the device. Look for signs of damage, cracking, colour changes, etc. to the medical
device during and after the application. In case of doubt, contact your local (Nutricia) sales representative.
• Avoid harsh or agressive cleaners/ disinfectants such as acetone, benzene or iodine.
• Do not sterilize the pump by irradiation, EtO, steam or in an autoclave.
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm soapy water, a 5% bleach
solution in water, a multipurpose disinfectant cleaner or a soft cloth.
• The Flocare
Infinity
pump may be rinsed by holding under a stream of warm, clean water. Do not submerge the pump!
®
• Always maintain the rollers on the rotor in a clean state to ensure they spin freely.
• The charger normally does not require cleaning. When desired, a dry or slightly damp cloth may be
used to clean the outside surface of the charger. Make sure the charger is disconnected from the wall outlet.
• Pump disposal: At the end of their service lifetime , the pump and its electrical accessories (AC Adapter Charger and Data
Download Accessory Cable), should be disposed of according to local standards and regulations governing the disposal of
electronic waste (e-waste). Other accessories can be disposed of or recycled as standard non-hazardous waste.
In case any of the problems listed on the problem solver chart occur, the pump delivers an audible and visual alarm and stops working.
The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "START/STOP" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "START/STOP".
The problem solver chart gives a clear explanation of the alarms.
403260087_r1.indd 92
403260087_r1.indd 92
PDMS Cable (corporate code 35776) enables connection to an external PDMS system. Follow the
Infinity
PDMS Cable to connect the pump.
®
MAINTENANCE AND CLEANING
ALARM FUNCTIONS AND SAFETY FEATURES
-92-