Télécharger Imprimer la page

Nutricia flocare Infinity Mode D'emploi page 95

Masquer les pouces Voir aussi pour flocare Infinity:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 49
Using methods other than those described may cause the pump to function incorrectly.
Condition
NO SET
• The set is not fitted or wrongly fitted in the
• The pressure sensor area is dirty ( Fig 1,
PUSH STRT
• The pump has been untouched in hold
END OF DOSE
• The pump administered the installed dose
PROG
• No flow rate is installed.
DOOR
• The door is not correctly closed.
AIR
• The pump detected air in the pump set.
• The air sensor area is dirty( Fig 1, position
• The feeding set is not inserted correctly.
BATT
• The battery capacity is too low.
the pump
remains working
Battery, "E" and "F"
• Battery failure.
flashing
FILL SET
• The pump is priming the set.
OCC IN
• The pump detected an upstream occlu-
• The pressure sensor area is dirty ( Fig 1,
• The pump detected a
OCC OUT
• The pressure sensor area is dirty
403260087_r1.indd 93
403260087_r1.indd 93
PROBLEM SOLVER CHART
Always follow the instructions below in case a problem occurs.
Cause
pump.
position 5 & 6)
mode for 3 minutes or more.
= volume.
Flow rate = 0 ml/h
8).
The pump is not able to deliver
highly viscous fluids at high flow rates with
the present charge level
of the battery.
sion between the pump and the feeding
bag.
position 5 & 6).
downstream occlusion between the pump
and the patient.
(Fig 1, position 5 & 6).
Correction
• Stop the alarm by pressing "START/STOP"
• Insert the Flocare
Infinity
feeding set into the pump as indicated on
®
the blister packaging (or refer to Fig 5a - 5d) and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the pump and restart
the pump.
• Stop the alarm and prolong the hold mode with another 3 minutes
by pressing "START/STOP" key.
• Program the pump and start it by pressing the "START/STOP".
• The programmed dose/volume has been administered
• Press start/stop to switch off the alarm
• Press INFO and then CLR to reset the administered volume (to reset
the flow rate and dose as well, press the CLR button for 3 seconds)
• Make sure the correct flow rate is programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Make sure there is no excessive foam in the reservoir and ensure
there is enough feed in the reservoir for the set therapy.
• If the feeding reservoir is empty, replace the reservoir and continue
feeding. If necessary, prime the set (FILL SET).
• Make sure the air sensor is clean.
• Make sure the feeding set is properly inserted in the pump. (Fig 5a -
5d)
• Connect the charger to the pump and mains and charge the pump
for approx. 6 hours. During charging the pump can be used.
• Turn the pump "off", contact your healthcare provider or refer the
pump to service.
• Press the "FILL SET" key another time to stop the pump and bring it
back in the hold mode.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the permeability
by flushing the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump and restart
the pump.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the permeability
by flushing the line.
• Check the permeability of the feeding tube.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump and restart
the pump.
-93-
EN
18/04/23 10:27
18/04/23 10:27

Publicité

loading

Produits Connexes pour Nutricia flocare Infinity