Télécharger Imprimer la page

DeVilbiss 525 Serie Manuel D'instructions page 9

Masquer les pouces Voir aussi pour 525 Serie:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 19
Catalog Number
Equipment Class and Type
Approval Body and Safety Standard
EMC Compliance To
*NOTE– The OSD performance at 41˚F (5˚C) to 104˚F (40˚C), 95% R.H. through voltage range on the 525DS verified at 670m.
**The maximum recommended flow is 3 LPM when an oxygen bottle is being filled with oxygen from the auxiliary oxygen port.
Specifications subject to change without notice.
DEVILBISS GUIDANCE AND MANUFACTURER'S DECLARATION
WARNING
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the Electromagnetic
Compatibility [EMC] information provided in the accompanying documents.
Portable and Mobile RF Communications Equipment can affect Medical Electrical Equipment.
The equipment or system should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the equipment or system
should be observed to verify normal operation in the configuration in which it will be used.
NOTE– The EMC tables and other guidelines provide information to the customer or user that is essential in determining the suitability of the Equipment or System for the
Electromagnetic Environment of use, and in managing the Electromagnetic Environment of use to permit the Equipment or System to perform its intended use without
disturbing other Equipment and Systems or non-medical electrical equipment.
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an
environment .
Emissions Test
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker
emissions
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The customer or the user of this device should assure that it is used in such an
environment .
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Radiated RF IEC 61000-4-3
Conducted RF IEC 61000-4-6
Electrical fast transient
IEC 61000-4-4
Surge IEC 61000-4-5
Power frequency magnetic field
IEC 61000-4-8
Voltage dips, short interrupts and
voltage variations on power supply
input lines IEC 61000-4-11
SE-525K3
525DS, 525DS-Q
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IPX0
TUV
IEC 60601-1 3rd Edition
EN60601-1-2
Compliance
Electromagnetic Environment – Guidance
This device uses RF energy only for its internal function . Therefore, its RF emissions are very low and
Group 1
are not likely to cause any interference in nearby electronic equipment .
Class B
Class B
This device is suitable for use in all establishments including domestic and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes .
Complies
IEC 60601 Test
Compliance
Level
Level
±6kV contact
Complies
±8kV air
3 V/m 80MHz to 2 .5GHz
Complies
3 Vrms 150kHz to 80MHz
Complies
±2kV power line
Complies
±1kV I/O lines
±1kV differential
Complies
±2kV common
3 A/m
Complies
>95% dip 0.5 cycle
60% dip 5 cycles
Complies
70% dip 25 cycles
95% dip 5 secs.
525KS, 525KS-LT
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IPX0
TUV approved for 50 Hz only to
EN 60601-1:1990+A1+A2+A13
ISO 8359: 1996/Amd .1:2012
EN60601-1-2
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic
material, the relative humidity should be at least 30%.
Field strengths outside the shielded location from fixed RF transmitters, as
determined by an electromagnetic site survey, should be less than 3 V/m .
Interference may occur in the vicinity of equipment marked with the following
symbol:
Mains power quality should be that of a typical commercial or hospital environment .
Power frequency magnetic fields should be at levels characteristic of a typical
location in a typical commercial or hospital environment .
Mains power quality should be that of a typical commercial or hospital environment .
If the user of this device requires continued operation during power mains
interruptions, it is recommended that the device be powered from an uninterruptible
power supply or battery .
525PS
Class II Equipment Double Insulated;
Type B Applied Part
Ordinary Equipment, IPX0
TUV
EN 60601-1:1990+A1+A2+A13
ISO 8359: 1996/Amd .1:2012
EN60601-1-2
EN - 9

Publicité

loading