Maintenance

9.1

Sensor exchange

Sensor cap
Marking on the
sensor
Sensor holder
10

Accessoires

The service life of the sensor is about 1 year in
normal use. It is shortened with constant use
when the refrigerant concentration is high. Proper
maintenance can extend the life of the sensor.
Switch off the unit.
Unscrew the sensor protection cap and wash
and dry it.
Blow impurities off the sensor with dry air.
Important: Never expose the sensor directly to
compressed air!
Store the unit in a dry and clean place and re-
move the batteries if you are not going to use
it for a long time.
If the sensor is defective, error message 2E ap-
pears in the display. In this case, replace the sen-
sor as follows:
Unscrew the sensor cap.
Note the exact position of the sensor and pull
the sensor out of its holder by hand.
Place the new sensor in the holder. Make sure
that the marking on the sensor housing is po-
sitioned correctly, see the adjacent figure.
Also make sure that all the sensor legs are
properly seated in the holder.
Replacement Sensor Wöhler RL 200:
Art. No.13028
Maintenancet