Télécharger Imprimer la page

Vega FIBERTRAC 31 Mise En Service page 81

Masquer les pouces Voir aussi pour FIBERTRAC 31:

Publicité

FF descriptor
ERROR_TEXT
PMT_VOLTAGE_AC-
TUAL
STANDARDIZATION_
FACTOR
SERIALNUMBER
NAMUR_STATE
NULL_COUNT_RATE
COUNT_RATE_PMT
ADJ_DENSITY_ABS_
COEFF
DEV_SW_VER_ASCII
POINT_LEVEL_AD-
JUST_MODE
RELAY_VALUE_SEL
DIGITAL_IN
DIGITAL_IN_BOOL
ANALOG_IN
FF_CHANNEL_AVAI-
LABLE
FF_CHANNEL_USED
Mapping of Process Value Status
Hex
Quality
0x00
Bad
0x01
Bad
0x0C
Bad
0x10
Bad
0x1C
Bad
0x20
Bad
0x44
Uncertain
FIBERTRAC 31 • Foundation Fieldbus
Rel. In-
Description
dex
45
Error text
46
Current photomultiplier voltage
47
Factor for the real value correction
48
Serial number
49
NAMUR state
50
Zero count rate
51
Pulse rate photomultiplier (raw values)
52
Process absorption coefficient
53
Software version
54
Point level adjustment mode
55
Relay basic value
56
Frequency of digital input
57
State of digital input
58
Current on analog input
59
Available channels
60
Used channels
Sub-Status
Non-specific
Configuration Error
Device Failure
Sensor Failure
Out of Service
Transducer in MAN
Last Usable Value
Condition
Unexpected error
It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,
117, 120
Five or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,
86, 121, 122, 141
On user error codes: 40, 41, 53, 68, 123, 124, 125
Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application
Transducer block is in mode "Manual"
Three or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
13 Annexe
Unit
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
81

Publicité

loading