III. Conditions of use
EN
1. Indications for use
EOlife X is a radio equipment intended to the training of users on a mannikin to the manual ventilation during
cardiopulmonary resuscitation (CPR) .
2. Patients
EOlife X is indicated to be used on:
- adult simulated patient between 4'7" (140 cm) and 6'7" (200 cm)
- pediatric simulated patient between 1 year (above 2'5" = 75 cm) and 12 years (below 4'7" = 140 cm)
3. Environment of use
EOlife X is intended to be used in training session .
4. Users
EOlife X is intended to be used by healthcare professionals and rescuers .
5. Contraindications
EOlife X is not approved for the following uses:
- In a hyperbaric chamber
- Near magnetic resonance imaging machines
IV. Safety
Precautions
o Before using the product, read these instructions carefully . Failure to follow the instructions can lead to improper use
and expose the user or persons nearby to risks .
Use the device as intended (see "III . Conditions of use") .
Comply with the contraindications (see "III Conditions of use") .
Carefully read the safety instructions in these instructions for use .
Comply with all the sections in these instructions .
o The use of defective devices may put the user at risk .
Only use the EOlife X device and its accessories if they do not show any visible damage .
o The EOlife X device, its battery, its charger and the FlowSense X sensor must be stored and used as described in the
instructions for use (see "X . Technical data") .
o The device can be safely used outdoors, as it is rainproof, splashproof and cannot be penetrated by solid objects larger
than 1 mm (IP44 rating) . However, it should be handled with care and kept away from powerful water jets and dust .
o Always check the battery level before using EOlife X . We advise recharging the battery after each use . Change the battery
if the charge is below 20% . We also advise having a spare operational battery in case this is needed .
o Using a battery with a low charge at temperatures below 0°C greatly reduces the run time and can cause EOlife X to stop
functioning after a short time . If the device is used at low temperatures, the battery should be fully charged .
o Do not remove the battery when EOlife X is switched on .