Télécharger Imprimer la page

ARCHEON EOlife X Instructions D'utilisation page 14

Publicité

Les langues disponibles

Les langues disponibles

8. Real-time indication of how to ventilate:
EN
Volume < Ideal volume .
When the user ventilates, the bar graph shows in real time the quantity of air/oxygen that must be insufflated
based on ERC and AHA guidelines .
In continuous resuscitation mode, EOlife X generates a visual countdown to show the user when to ventilate
to deliver the 'target frequency' breaths per minute . When the green lung is displayed, the user must
ventilate .
o Display in 30:2 and 15:2 mode:
In 30:2 mode, the countdown function is disabled as the person ventilating must wait for 30 chest
compressions before delivering 2 ventilations .
In 15:2 mode, the countdown function is disabled as the person ventilating must wait for 15 chest
compressions before delivering 2 ventilations .
In this mode the ventilation frequency depends on the person (or machine) performing the chest
compressions .
The green lung is shown permanently to indicate that the device is waiting to detect a ventilation to display
the bar graph .
14
A bar graph shows whether the volume insufflated/exhaled is adequate:
Orange:
Display in continuous mode:
Green:
Ideal volume .
Red:
Volume > Ideal volume .

Publicité

loading