Connect the ventilation interface (mask or
endotracheal tube) to the compatible end of the
FlowSense X sensor . The sensor ends are shaped
differently to ensure that the interface cannot be
EN
connected to the wrong end .
Connect the bag to the other end of the FlowSense
X sensor . The device must be assembled as shown
here:
Disassembly
To remove the accessories from the product, take the FlowSense X sensor out of the ventilation interface (mask or
endotracheal tube) by gently pulling vertically on the FlowSense X sensor . Then remove the ventilation bag at the other
end in the same way .
b) EOlife X blind ventilation training mode
In blind ventilation mode, EOlife X calculates the volumes insufflated and expired, the ventilation frequency and
leakage, estimates the tidal volumes, and records the data . After recording the ventilation session, EOlife X provides a
statistical report on the quality of the ventilation performed .
c) EOlife X monitored ventilation training mode
In monitored ventilation mode, EOlife X gives the user real-time feedback on the quality of the ventilation performed
based on a cyclic analysis of inspiratory and expiratory flow rates . It must therefore be positioned proximally to be able
to measure both the inspiratory and expiratory flow rates, calculate the volumes insufflated and expired, the ventilation
frequency and leakage, estimate the tidal volumes, and generate user information signals and visual signals if the
ventilation parameters are unacceptable .
EOlife X also uses algorithms to filter out small airflows generated by chest compressions so as not to display
anomalous volume and ventilation frequency values and to avoid measurement artefacts . During ventilation, be sure to
generate enough flow to be detected by the device .
Warning
When ventilating with a mask, ensure it is correctly held on the manikin's face during insufflation and for the entire
duration of expiration so that EOlife X can measure the volume expired .
12
Medium Continuous Rotate
55%
Recording session
REC
STOP
Blind ventilation
training screen
Recording indicator
Click on stop to
stop recording