9. Visual signal display area :
"Good ventilation"
"Insufficient volume"
"Excessive volume"
"Low frequency"
"High frequency"
"Important leakage"
Note
See details of visual signals in "VIII . 1 c) Visual signals" .
d) Visual signals conditions
Visual messages related to ventilation parameters
Visual messages may appear during ventilation when the ventilation parameters exceed the threshold values and are
persistently outside the ranges given below .
The screen shows the visual messages in the visual signal display area (lower part of the screen) as a message .
The visual messages concern:
The tidal volume (Vt):
Visual signal
conditions
Insufficient volume
all the ventilation parameters calculated are within the appropriate range of
values as defined in the international guidelines .
the volume of air insufflated to the patient is significantly lower than the
values given in the international guidelines .
the volume of air insufflated to the patient is significantly higher than the
values given in the international guidelines .
the ventilation frequency is much significantly than the values given in the
international guidelines .
the ventilation frequency is significantly higher than the values given in the
international guidelines .
the difference calculated between the volume of air/oxygen insufflated
and the volume exhaled is too great . This may be because the mask is not
positioned correctly, due to improper intubation (oesophageal intubation) or
because the ventilation circuit is not properly sealed .
Cause
Patient height
Tidal volume value
Small
< 225 mL
Medium
< 285 mL
Tall
< 350 mL
1-2 years
< 45 mL
2-3 years
< 55 mL
3-4 years
< 70 mL
5-6 years
< 90 mL
7-9 years
< 115 mL
10-11 years
< 140 mL
Solution
Insufflate a volume of air within the
appropriate range as defined in the
international guidelines