Quick Start Guide
3.1 Electrical warnings
To prevent electrical shocks or damage to the instrument, do
not connect more than 30V CAT I between the terminals, or
between the terminals and the ground (earth).
External circuits should have appropriate insulation to the
mains.
To prevent electrical shocks, use only the GE specified AC
probe (Part: IO620-AC) to measure AC voltages that are more
than 20 Vrms. Do not connect more than 300V CAT II between
the IO620-AC leads, or between the leads and the ground
(earth). Attach it to the specified connections only.
This instrument uses a Lithium-Polymer (Li-Polymer) battery
pack. To prevent an explosion or fire, do not short circuit, do
not disassemble, and keep it safe from damage.
To prevent an explosion or fire, use only the GE specified
battery (Part: 191-356), power supply (Part: 191-339) and
battery charger (Part: IO620-CHARGER).
To prevent battery leakage or heat generation, only use the
battery charger and power supply in the temperature range 0
to 40°C (32 to 104°F).
The power supply input range is 100 – 240Vac, 50 to 60Hz,
250mA, installation category CAT II.
Position the power supply so not to obstruct the supply
disconnecting device.
To make sure the display shows the correct data, disconnect
the test leads before power is set to on or change to another
measure or source function.
To prevent a dangerous release of pressure, make sure that all
the related pipes, hoses and equipment have the correct
pressure rating, are safe to use and are correctly attached.
3 May 2013
WARNING
K0542 Issue 2
Druck DPI 620 Genii