Immunity test
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
Portable and mobile RF communications equipment should not be used closer to any part of
VisCalor® Dispenser, including cables, than the recommended separation distance calculated from the
Conducted RF IEC
61000-4-6
Radiated RF
IEC 61000-4-3
32
IEC 60601 test level
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for 5 sec
3 A/m
equation applicable to the frequency of the transmitter.
3 Vrms
150kHz to 80 MHz
3V/m
80MHz to 2.5 GHz
Compliance level
70% UT
(30% dip in UT) for 25
cycles
<5% UT
(>95% dip in UT) for 5 sec
3 A/m
Recommended separation distance
3V
10V/m
Dispenser
Electromagnetic
environment - guidance
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
dental environment.
VisCalor® Dispenser
d = 0.35 √P
d = 0.35 √P 80 MHz to
800 MHz
d = 0.70 √P 80 MHz to
2.5 MHz