INSTRUCTIONS FOR USE
• PPE must be worn during manual decontamination to prevent the risk of infection.
• Deviating from the specified cleaning instructions could cause a biological hazard, especially in
multi-user environments, and adversely affect the life and efficiency of the product.
• Regular cleaning can help prolong the lifespan of the powerchair.
Caution
The surface of the powerchair frame has been treated with a protective coating to allow for easy removal of dirt. Use
a damp cloth with a mild soap to clean the frame and then dry thoroughly.
11. MAINTENANCE & SERVICING
• Your powerchair should be serviced every 12 months, take your powerchair to a service engineer
for inspection and maintenance who are authorised on behalf of Drive DeVilbiss Healthcare Ltd.
Warning
• Always disconnect the powerchair from the mains power supply and remove the battery prior to
performing any maintenance procedures (where viable).
• Never attempt to open the enclosures, re-wire any components or replace internal battery com-
ponents. Modification of the powerchair is not allowed without the permission of Drive DeVilbiss
Healthcare ltd and electrical system components are only to be replaced by authorised service
personnel.
• No maintenance or servicing should be conducted while the device is in use – risk of electric shock,
entrapment, etc. If not possible due to the occupant's mobility, a risk assessment should be carried
out, and if deemed safe to proceed, care should be taken to avoid contact with the occ pant when
working on electrical items.
• Failure to carry out the following checks at the stated frequencies could negatively influence the
essential performance of the powerchair and as a result put individuals at risk.
• Allow all components to cool before performing maintenance. Parts of the motor system can gene-
rate heat while driving.
• All maintenance should be conducted by a competent person. Disassembling the controller, motor,
or charger by anyone other than an approved service engineer is prohibited and voids any appli-
cable warranty. For any maintenance concerns, contact Drive DeVilbiss Healthcare Germany.
Caution
11.1 General Guidelines
Routine maintenance is required to ensure the maximum use of your powerchair. While some of the maintenance can
be done by yourself, you may need assistance from an authorised service engineer. If you have any doubts, contact
Drive DeVilbiss Healthcare Ltd.
If there are any signs of damage, or the powerchair is not performing as it should, withdraw it from service until the
powerchair has been repaired and is fit for use. Preventative maintenance is key to keeping the AirFold Powerchair
in prime operating condition. Follow the Maintenance Schedule at the end of this manual to periodically inspect the
AirFold Powerchair for serviceable items.
• Avoid abuse to the joystick and control functions
• Avoid prolonged exposure to extreme heat or cold
• Keep the powerchair clean and free from moisture
• Never use a conditioner on the tread of the wheels
AirFold Powerchair
EN
59